Tips for dimension measurement using transmitted light

When performing dimension measurement using transmitted light, the appearance can vary depending on the “type of transmitted light” and the “brightness” used.

 

This article introduces the “recommended lighting” and “brightness” for dimension measurement using transmitted light, so please read it through to the end.

 

 

・Use “parallel transmitted light”

There are several types of transmitted lighting devices.

 

 ・Diffused transmitted light (general type of transmitted light)

 ・Parallel transmitted light (compact transmitted light suitable for dimension measurement)

 ・Telecentric light (high-precision, large-scale light used in projectors and similar devices)

 

Of course, telecentric light is the most suitable for dimension measurement.

However, it is large in size and significantly more expensive (it is integrated into high-precision projectors and dimension measuring instruments).

 

For general use, we recommend parallel transmitted light.

 

 

The lighting method is particularly important when measuring samples with significant thickness.

(For thin samples, the difference is less noticeable.)

 

透過照明で寸法計測01  

We measured the Mitutoyo block gauge with a width of 9 mm.

 

<Diffused transmitted light>   <Parallel transmitted light>
透過照明で寸法計測02   透過照明で寸法計測03

*Each was measured twice consecutively.

 

For items such as glass scales, either lighting method is suitable.

However, when the specimen has thickness, we recommend parallel transmitted light.

 

 

 

 

・Adjust the brightness to a “darker” setting

Generally, adjust to a “darker” setting.

Increasing the illuminance causes the white areas (bright spots) to expand.

 

I measured a glass scale with a radius of 2.5 mm using our non-calibrated microscope.

(When the measured area is white, increasing the brightness causes it to appear larger.)
* The “Edge automatic detection mode (A)” was set to eliminate variation due to the operator.

 

<When adjusted to a brighter setting>   When adjusted to a darker setting
透過照明で寸法計測04   透過照明で寸法計測05
⇓ Zoom in on the area within the red frame above   ⇓ Zoom in on the area within the red frame above
透過照明で寸法計測06   透過照明で寸法計測07

 

 

With our non-calibrated microscope, it is recommended to adjust to a slightly darker setting, within the range where the automatic edge detection mode (A) functions stably.

 

 

 

 

 

Dimension measurement (auto-calibration) microscope (medium magnification)

CT200HD

 

寸法測定(自動校正)マイクロスコープ(高倍率) CT200HD-H  

Dimension measurement (auto-calibration) microscope (high magnification)

CT200HD-H