Method for Using the Ultra-High Magnification Microscope (NSH500CSU) with Transmitted Illumination

The NSH500CSU comes with a standard simple XY table, but it does not support transmitted illumination. Here, we will introduce a method to attach transmitted illumination.

**Pattern 1**
Attach rubber feet to the RD-95T and place it on the standard included XY stage (TK100).

 

RD-95Tにゴム足を取付け、標準付属のXYステージ(TK100)に載せる方法

 

This method allows for the easy introduction of transmitted illumination without the need for modifications, but the size of the specimen is limited to the dimensions of the RD-95T (φ95).

While the rubber feet prevent slipping, the lighting unit may shift due to impacts such as accidental hand contact.

**Pattern 2**
Replace the rotating simple XY stage, remove the observation plate, and insert the RD-95T.

 

回転式簡易XYステージに変更し、観察板を外しRD-95Tをはめ込む方法

 

The RD-95T can be fitted into the XY table, providing a certain degree of stability. However, since the cable needs to be routed outside, it is necessary to drill a hole of approximately φ10 in the base. Our company can perform this drilling free of charge prior to shipment.